Investigations of a method for determining pumping speed and sorption capacity of nonevaporable getters based on in situ calibrated throughputErjavec, Bojan, 1950-2023 ; Šetina, JanezThis paper presents a gas-sorption method, based on an in situ calibrated throughput, for characterizing nonevaporable getters (NEGs) of Ti- and Zr-based alloys. The main characteristics of the ... method are the adjustment of a constant injected gas flow, the prompt calibration of a throughput, and the measurement of an increasing sorption pressure. It is essential that inert vacuum gauges be selected for measurements of total pressure to prevent contamination of a test gas and poisoning of an NEG under test. A capacitance diaphragm gauge is used for preparation of the injected gas flow by filling the gas reservoir with a test gas and adjusting the conductance of a variable leak valve. A spinning rotor gauge (SRG) is used for in situ calibration of the throughput by means of an integrated gas-flow calibration facility and the rate-of-pressure-rise method, and by continuous measurement of the sorption pressure in a test chamber. A getter pumping speed and a sorbed quantity are calculated while considering the trhoughput, the sorption pressure, and the sorption time. The throughput must be selected in accondance with both the sorption characteristics of a particular Neg TYPE and the measurement capabilities of the SRG. At room temperature, porous thick-film NEGs and structured thin-film NEGs exhibit an initial getter pumping speed on the order of 1 and 0.1 1 s[sup]{-1} cm[sup]{-2}, respectively. The corresponding sorption capacity of a few times 10[sup{-4} mbar 1 cm[sup]{-2} is comparable for all chemically active gases except for H[sub]2, which amounts to a few times 100 x 10[sup]{-3} mbar 1 cm[sup]{-2} in the case of porous NEGs and ~ 1 x 10[sup]{-3} mbar 1 cm[sup]{-2} in the case of structured NEGs. Detailed analysis of uncertainties of SRG pressure measurements below 1 x 10[sup]{-3} mbar gives an expanded uncertainty of the getter pumping speed and the sorbed quantity of 1.9% and 3.5%, respectively. While temporally interrupting the throughput, the setup enables discrete measurements of the background pressure, which increases because of the acumulation of nongetterable gases. After completion of the gas-sorption test, a qualitative determination of the accumulated inert gas composition is performed by the gas-burst method using a quadrupole mass spectrometer (QMS) mounted on a vacuum system. The quantitative analysis is enabled by the use of a QMS calibrated in situ by known amounts of a certain gas such as H[sub]2, CH[sub]4, CO, Ar and Kr.Vir: Journal of vacuum science & technology. A, Vacuum, surfaces, and films. - ISSN 0734-2101 (Vol. 29, no. 5, sept./oct. 2011, str. 051602-1/051602-10)Vrsta gradiva - članek, sestavni del ; neleposlovje za odrasleLeto - 2011Jezik - angleškiCOBISS.SI-ID - 874154
vir: Journal of vacuum science & technology. A, Vacuum, surfaces, and films. - ISSN 0734-2101 (Vol. 29, no. 5, sept./oct. 2011, str. 051602-1/051602-10)
Vnos na polico
Trajna povezava
- URL:
Faktor vpliva
Dostop do baze podatkov JCR je dovoljen samo uporabnikom iz Slovenije. Vaš trenutni IP-naslov ni na seznamu dovoljenih za dostop, zato je potrebna avtentikacija z ustreznim računom AAI.
| Leto | Faktor vpliva | Izdaja | Kategorija | Razvrstitev | ||||
|---|---|---|---|---|---|---|---|---|
| JCR | SNIP | JCR | SNIP | JCR | SNIP | JCR | SNIP | |
Faktor vpliva
Baze podatkov, v katerih je revija indeksirana
| Ime baze podatkov | Področje | Leto |
|---|
| Povezave do osebnih bibliografij avtorjev | Povezave do podatkov o raziskovalcih v sistemu SICRIS |
|---|---|
| Erjavec, Bojan, 1950-2023 | 05673 |
| Šetina, Janez | 04254 |
Vir: Osebne bibliografije
in: SICRIS
Izberite prevzemno mesto:
Prevzem gradiva po pošti
Naslov za dostavo:
Med podatki člana manjka naslov.
Storitev za pridobivanje naslova trenutno ni dostopna, prosimo, poskusite še enkrat.
S klikom na gumb "V redu" boste potrdili zgoraj izbrano prevzemno mesto in dokončali postopek rezervacije.
S klikom na gumb "V redu" boste potrdili zgoraj izbrano prevzemno mesto in naslov za dostavo ter dokončali postopek rezervacije.
S klikom na gumb "V redu" boste potrdili zgoraj izbrani naslov za dostavo in dokončali postopek rezervacije.
Obvestilo
Trenutno je storitev za avtomatsko prijavo in rezervacijo nedostopna. Gradivo lahko rezervirate sami na portalu Biblos ali ponovno poskusite tukaj kasneje.
Gesla v Splošnem geslovniku COBISS
Izbira mesta prevzema
Gradivo iz matične enote je brezplačno. Če je gradivo na mesto prevzema dostavljeno iz drugih enot, lahko knjižnica to storitev zaračuna.
| Mesto prevzema | Status gradiva | Rezervacija |
|---|
Rezervacija v teku
Prosimo, počakajte trenutek.
Rezervacija je uspela.
Rezervacija ni uspela.
Rezervacija...
Članska izkaznica:
Mesto prevzema: